Specifications
pH input
Input for digital sensors (RS485)
Display range
pH value
–2.00 ... 16.00
ORP
–1999 ... 1999 mV
Temperature
–20.0 ... +200.0 °C (-4 ... + 392 °F)
pH sensor standardization
*
pH calibration
Operating modes
BUF
Calibration with Calimatic
automatic buffer recognition
MAN
Manual calibration with
entry of individual buffer values
DAT
Data entry of pre-measured electrodes
Product calibration
Calimatic buffer sets
*
-01- Mettler-Toledo
2.00/4.01/7.00/9.21
-02- Knick CaliMat
2.00/4.00/7.00/9.00/12.00
-03- Ciba (94)
2.06/4.00/7.00/10.00
-04- NIST technical
1.68/4.00/7.00/10.01/12.46
-05- NIST standard
1.679/4.006/6.865/9.180
-06- HACH
4.01/7.00/10.01
-07- WTW techn. buffers 2.00/4.01/7.00/10.00
-08- Hamilton
4.01/7.00/10.01/12.00
-09- Reagecon
2.00/4.00/7.00/9.00/12.00
-10- DIN 19267
1.09/4.65/6.79/9.23/12.75
-U1-
Specifiable buffer set with 2 buffer solutions
Zero adjustment
± 200 mV (for ISFET)
Max. calibration range
Asymmetry potential
±60 mV
Slope
80 ... 103 % (47.5 ... 61 mV/pH)
(possibly restricting notes from Sensoface)
ORP sensor standardization
*
ORP calibration (zero adjustment)
Max. calibration range
–700 ... +700 ΔmV
Adaptive cal timer
*
Interval 0000 ... 9999 h (Patent DE 101 41 408)
Sensocheck
Automatic monitoring of glass electrode (can be disabled)
Delay
Approx. 30 s
Sensoface
Provides information on the sensor condition,
evaluation of zero/slope, response time, calibration interval,
wear, Sensocheck, can be switched off