71
Error messages can be signaled by a 22 mA output current (see Error Messages and
Configuration of Output 1/Output 2).
The alarm delay time delays the color change of the display backlighting to red
and the 22 mA signal (if configured).
Configuration
Menu item
Action
Choices
Alarm delay
Enter value using
keys.
Press enter to confirm.
0...600 SEC
(010 SEC)
Sensocheck
Select Sensocheck
(continuous monitoring
of glass and reference
electrode)
Select ON or OFF using
keys.
Press enter to confirm.
(At the same time,
Sensoface is activated.
With OFF, Sensoface is
also switched off.)
ON/OFF
5