110
Specifications
Diagnostics functions
Calibration data
Calibration date, zero, slope, response time
Device self-test
Display test, automatic memory test (RAM, FLASH, EEPROM)
Logbook
100 events with date and time
Extended logbook (TAN)
Audit Trail: 200 events with date and time
Service functions
Sensor monitor
Display of direct sensor signals
Current source
Current specifiable for output 1 and 2 (04.00 ... 22.00 mA)
Passcodes
Assigning passcodes for menu access
Factory setting
Resetting all parameters to factory setting
TAN
Enabling optionally available additional functions
Explosion protection
(A2**B/X)
See Control Drawing or www.knick.de
Data retention
Parameters, calibration data, logbook > 10 years (EEPROM)
EMC
EN 61326-1 (General Requirements)
Emitted interference
Class B (residential area)
Immunity to interference
Industry EN 61326-2-3
Nominal operating conditions
Ambient temperature
–20 ... +65 °C / -4 ... +149 °F
Transport/Storage temperature
–30 ... +70 °C / -22 ... +158 °F
Relative humidity
10 ... 95% not condensing
Supply voltage
14 ... 30 V