![Knick Stratos Pro A2 COND Series User Manual Download Page 76](http://html.mh-extra.com/html/knick/stratos-pro-a2-cond-series/stratos-pro-a2-cond-series_user-manual_1980189076.webp)
76
Measured values
pH value
pH voltage
Temperature
7.09
pH
49.2
mV
25.1
°C
Sensor data
Sensor type:
Manufacturer:
Order code:
Serial number:
KNICK
SE 533X/1-NMSN
1030550
Adjustment data
Date:
Slope:
Zero point:
06/27/2011 20:09:12
58.5 mV/pH
7.06 pH
StartCenter
Calibration
Table View
History
Statistics
pH Buffers
Sensor type:
Manufacturer:
pH (glass)
KNICK
Order code:
Serial number:
SE 533X/1-NMSN
1030550
Measuring point:
Tag number:
0
Change
Change
Show
pH (glass)
Digital Sensors
Stratos Pro can be operated with digital sensors. Due to the galvanic
isolation of Memosens sensors, earth or ground potentials have no
effect here. Therefore, a Solution Ground or measures for equipotential
bonding are not required.
Digital sensors can be calibrated and maintained in the lab.
This considerably simplifies on-site maintenance.
Memosens Sensors:
Calibration and Maintenance in the Lab
The “MemoSuite“ software allows calibrating Memosens sensors
under reproducible conditions at a PC in the lab. The sensor param-
eters are registered in a database. Documenting and archiving meet
the demands of FDA CFR 21 Part 11. Detailed reports can be output as
csv export for Excel.
Currently connected sensor:
Sensor type, manufacturer,
order code and serial number
Settings and
specifications
Function selection
(The selected function is
highlighted.)
Display size of measured values:
When the cursor moves over a measured value, it
changes to a magnifying glass, allowing to magnify
the measured-value display at a mouse click.
Parameters of currently
connected sensor
Last calibration
(adjustment)
pH value
7.09
pH
Measured values
pH voltage
49.2 mV
Summary of Contents for Stratos Pro A2 COND Series
Page 121: ...121 Buffer Tables 10 DIN 19267 buffers ...
Page 141: ...141 Index ...