3 Troubleshooting
Error messages
Range
limits
exceeded
If a measured value lies outside the ranges accepted by the
meter, an error message appears and the measured-value
display flashes.
ERROR 1
The measurement range was exceeded.
Possible causes:
• Sensor defective
• Break in sensor cable
• Wrong sensor connected
• Wrong cell constant entered
ERROR 3
The measured temperature is outside the ranges:
Conductivity:
nLF:
Salinity:
TDS:
–20 °C to +120 °C
0 °C to 120 °C
0 °C to 30 °C
10 °C to 40 °C
Possible causes:
• Temperature detector in the sensor defective
• Short circuit in temperature detector
• Wrong temperature detector connected
Note
When changing the conductivity sensor, note that the
temperature detector type (Pt 1000/NTC 30 kΩ ) is only
recognized when the meter is switched on with on/off.
Troubleshooting