Operation
Data memory
Up to one hundred measured values can be saved in the
data memory together with the temperature, date and
time. Storage is performed either manually or automati-
cally using the data logger. The currently measured value
is stored.
Write memory Press STO.
The currently measured value is shown in the display.
Select any memory location with
▲
and
▼
.
Press STO to store the measured value in the selected
memory location
After storing, the memory location number is automatical-
ly incremented and the meter returns to measuring mode.
Read memory Pressing RCL displays the last stored value.
Select any memory location using
▲
and
▼
.
Pressing RCL switches between the measured value and
the time/date of storage.
This allows, for example, searching for a value that was
stored at a certain time.
Pressing meas returns to the measuring mode.
Clear memory To clear the entire data memory, press STO to access the
memory mode and then press clock to access the data
logger mode.
Here, select Clear (
CLr
) using
▲
or
▼
.
+
By confirming this with STO, the entire data memory is
cleared.
If you do not want to clear the memory, press meas to
cancel.