Batteries for applications in hazardous locations
Batteries (3 x each)
Temp. class
Ambient temperature range
Duracell MN1500
T4
–10 °C ≤ Ta ≤ +40 °C
Energizer E91
T3
–10 °C ≤ Ta ≤ +50 °C
Power One 4106
T3
–10 °C ≤ Ta ≤ +50 °C
Panasonic Pro Power LR6
T3
–10 °C ≤ Ta ≤ +50 °C
The batteries listed in the table are used instead of the Varta batteries Type
4006, Type 8006 and Type 3706 listed in the EC-Type-Examination Certificate.
These batteries were tested according to IEC 60079-0:2007, IEC 60079-11:
2006 by a notified body and the test results were set out in an IECEx test
report.
Equipped with the batteries listed in the table, the Model 91. X Cond
portable conductivity meter may be used in hazardous areas.
Be sure to observe the temperature classes and ambient temperature ranges
listed in the table for the different battery types.