Operation
Automatic calibration (
AutCAL On
) with 0.1 molar KCl
solution, 0.01 molar KCl solution or entry of the tem-
perature-compensated conductivity of another known
calibration solution.
Automatic
switch-off
To protect the batteries, the meter switches off
automatically when not operated for a longer time.
You can select whether switch-off is to take place after
one hour or after twelve hours (default setting: 1 hour).
If the data logger is active and during remote interface
operation, the auto switch-off feature is disabled.
Interface
If the meter is controlled by a PC and interface conflicts
occur when the print key is pressed, you should deacti-
vate the print function
(Print OFF)
(default setting: Print On, 4,800 bauds).
The transmission speed can be set to 600, 1200, 2400,
4800 or 9600 bauds.
The transmission speed must correspond to that set in
the printer or PC.
Data format and protocol are permanently set to 7
bits, one stop bit, even parity and XON/XOFF protocol
(NAMUR NE28).
Temperature
display
The temperature can be displayed either in °C or °F
(default setting: °C).