Novar 1xxx
KMB systems
16
2. Installation
2.1 Physical
Instruments designed for a distribution board panel are mounted into a cutout of required size. The
instrument’s position must be fixed with locks.
The Novar1005D and theNovar1007D are designed for mounting on a DIN-35 bar.
Natural air circulation should be provided inside the distribution board cabinet, and in the instrument’s
neighbourhood, especially underneath the instrument, no other instrumentation that is source of heat
should be installed or the temperature value measured may be false.
2.2 Connection
To connect the controller there are connectors with screw-on terminals in the back wall. Signal pinout
on these connectors can be seen from the pictures below.
Examples of controller wiring are shown in a separate chapter.
Fig. 2.1 : Novar1114 controller – connectors
L1
L2
L3
N
1
2
3
4
5
6
7
8
9
10
11
12
13
14
ALARM
5A
(1A)
max.
10A
100÷275 VAC
k
l
L
N
RS 232/485
Rx
TR
A
B GND
Tx GND
Made in Czech Republic
/
/
NOVAR 1106 1114 / 232 485
Serial / vers.:
Product. date :
IP 4X
U 100 275 VAC, 7VA,
÷
÷
43 67 Hz
SUPPLY
LOAD
Maximum cross section area of connection wires is 2.5 square millimetres.
2.2.1 Power Supply
2.2.1.1 Standard Version Controllers
The controller requires supply voltage in the range as declared in technical specifications table for its
operation.
The supply voltage connects to terminals 3 (
L
) and 4 (
N
). In case of DC supply voltage the polarity of
connection is free. Power supply voltage needs to be externally protected ( see chapter
Protection
below ).
The 12xx line controllers have power supply terminals 3 (
L
) and 4 (
N
) internally connected to terminals
5 (
L
) and 6 (
N
) which can be used to connect the power supply voltage to measurement voltage input
(terminals 7 –
L
and 9 –
N/L
).
Power supply terminal 3 (
L
) is internally connected to the common pole of output relays. It is
necessary to dimension the power supply protection in consideration of output contactors’ power as
well.
Summary of Contents for NOVAR-1 Series
Page 66: ...Novar 1xxx KMB systems 66 Novar1106 installation...
Page 67: ...Novar 1xxx KMB systems 67 Novar1114 installation...
Page 68: ...Novar 1xxx KMB systems 68 Novar1206 installation low voltage measurement...
Page 69: ...Novar 1xxx KMB systems 69 Novar1214 installation high voltage measurement...
Page 70: ...Novar 1xxx KMB systems 70 Novar1114 S400 installation...
Page 73: ...Novar 1xxx KMB systems 73 Novar1414 installation low voltage measurement...