24. High Temp
Error
SVS2000 is sensing a temperature inside
enclosure that is above SvS2000’s specified
operating temperature
yes
1. Check temperature just outside the enclosure. If
temperature is greater than 122° F (50° C), application
does not meet the SvS2000’s specifications.
2. Check inside enclosure to see if a component failed
and is dissipating heat. If component failure is causing
error message you may notice an unusual smell, a hot
component, and/or a charred component. Contact KM.
3. If temperature inside and outside unit meet
specifications, temperature sensor on microprocessor/
Display PCB in SvS2000 may be damaged — contact
KM. CAUTIOn Turn off power to SVS2000 and let it
cool down.
hoT KeyS
Following is a summary of the keyboard functions (hot keys) used in run mode:
Switch from Run Mode to Function Mode (menu tree)
Press the Func Key
Switch from Run Mode to a particular menu item in Function Mode
Press the two-digit function number and then press the Enter Key.
Switch from Function Mode to Run Mode
Press the Gross or net Key
Scroll through the Run Mode display screens
(gross, net, total, and relay status)
Press the Up Arrow or Down Arrow Key
Switch from Gross to net Weight
Press the net Key
Switch from net to Gross Weight
Press the Gross Key.
Switch from gross weight numerical display to Bar graph display or
vice versa
Press the Enter Key while on the gross weight display to toggle
between numerical and bar graph display.
Zero Gross Weight
(refine the calibration when the vessel is empty) —
Press the Zero Key while on the gross weight display. When the
display requests verification that you want to zero the gross weight,
press the Enter Key.
Press any other key to abort the operation.
Zero Total Weight
Press the Zero Key while on the total weight display. When the
display requests verification that you want to zero the total weight,
press the Enter Key.
Press any other key to abort the operation.
Display number of Tares
Press the Enter Key while on the total weight display. The display
briefly shows the number of tares that have occurred since the total
weight was zeroed.
Tare Vessel
Press the Tare key for 1 second, until the display flashes Tare
Implemented.
Display SVS2000 Serial Address in hexadecimal form
Press the ‘-’ Key and then the ‘1’ Key.
Display Current SVS2000 Software Revision letter and Date of
Release
Press the ‘-’ Key and then the ‘2’ Key
Summary of Contents for SVS2000
Page 1: ...SVS2000 Installation Operation Manual IOM97 1129 01 Rev D...
Page 2: ......
Page 10: ...6 97 1129 01 Rev D www kistlermorse com...
Page 11: ...7 www kistlermorse com 97 1129 01 Rev D...
Page 12: ...8 97 1129 01 Rev D www kistlermorse com...
Page 13: ...9 www kistlermorse com 97 1129 01 Rev D...
Page 14: ...10 97 1129 01 Rev D www kistlermorse com...
Page 20: ...16 97 1129 01 Rev D www kistlermorse com Figure 3 Wiring Diagram Current Output Card venture...