Preact [31 through 38]
IOut 4/0 mA [42] and 20 mA [43]
DSP Step [73]
rate Threshld [81]
Zero Trak Win [83]
Drift Limit [84]
Example 2. Change Disp Units from lbs to tns and Disp Format from xxxxxx00 to xxxxx.x
record current parameter values:
Disp Format = xxxxxx00
Scale Factor Weight = 100,000 lbs
Relay1 Value = 90,000 lbs
(dummy zeroes underlined)
Change Disp Units and Disp Format.
As a result of the changes, the parameter values change to:
Scale Factor Weight = 100.0 tns
Relay1 Value = 90.0 tns
re-enter parameter values:
Scale Factor Weight =
100,000 lbs ÷ 2000 lbs/ton = 50.0 tns
Relay1 Value =
90,000 lbs ÷ 2000 lbs/ton = 45.0 tns
Example 3. Change Disp Units from lbs to % and Disp Format from xxxxxx00 to xxxxxx
When changing Disp Units to %, you must calculate the current parameter values in terms of the full scale
(100%) load:
new value (in %) = Old value ÷ Full Scale Load where:
Full scale load = vessel working capacity (from Application Data Form)
Current parameter values:
Disp Format = xxxxxx00
Scale Factor Weight = 100,000 lbs
Relay1 Value = 90,000 lbs (dummy zeroes underlined) Change Disp Units and Disp Format. As a result of the
changes, the parameter values change to:
Scale Factor Weight = 1000 %
Relay1 Value = 900 %
The vessel working capacity is 150,000 lbs.
re-enter parameter values:
Scale Factor Weight =
100,000 lbs ÷ 150,000 lbs = 75%
Relay1 Value =
90,000 lbs ÷ 150,000 lbs = 60%
Summary of Contents for SVS2000
Page 1: ...SVS2000 Installation Operation Manual IOM97 1129 01 Rev D...
Page 2: ......
Page 10: ...6 97 1129 01 Rev D www kistlermorse com...
Page 11: ...7 www kistlermorse com 97 1129 01 Rev D...
Page 12: ...8 97 1129 01 Rev D www kistlermorse com...
Page 13: ...9 www kistlermorse com 97 1129 01 Rev D...
Page 14: ...10 97 1129 01 Rev D www kistlermorse com...
Page 20: ...16 97 1129 01 Rev D www kistlermorse com Figure 3 Wiring Diagram Current Output Card venture...