![KLS Martin LIMAX 120 Service Manual Download Page 81](http://html1.mh-extra.com/html/kls-martin/limax-120/limax-120_service-manual_1979001081.webp)
Service manual LIMAX
120
Service manual version 1.00
page 78
Status-Code
Error
description
First step done
by user
Second step
done by user
Service
operation done
by technician
2084
Grave error:
Photosensor 2
value low
(pulsphase fall)
Press down
STANDBY
call service
Replace
controller PCB
2085
Grave error:
Photosensor 2
value high
(pulsphase fall)
Press down
STANDBY
call service
Replace
controller PCB
2086
Warning:
Photosensor 1
deactivated
-
-
-
2087
Warning:
Photosensor 2
deactivated
-
-
-
2088
Grave error:
Controller
system activates
undefined
Pulsmode
Press down
STANBY
Call service
Replace
controller PCB or
power supply
unit
2089
Grave error:
Controller
system changes
undefined laser
parameter
Press down
STANBY
Call service
Replace
controller PCB or
power supply
unit
2101
Grave error:
Controller
system activates
SWAP for
programs
without SWAP
component
Press down
STANDBY
Call service
Replace
controller PCB or
power supply
unit
2102
Grave error:
Controller
system chooses
undefined SWAP
component
Press down
STANDBY
Call service
Replace
controller PCB or
power supply
unit