Service manual LIMAX
120
Service manual version 1.00
page 68
Status-Code
Error
description
First step done
by user
Second step
done by user
Service
operation done
by technician
1009
Critical error:
MOS field effect
transistor failure
Power off and on
call service
Test/ replace
power supply
unit
1011
Critical error:
Atmel CPU
power failure
Power off and on
call service
Replace
controller PCB
1012
Critical error:
Atmel CPU self-
test failure
Power off and on
call service
Replace
controller PCB
1013
Critical error:
Atmel CPU not
responding
Power off and on
call service
Replace
controller PCB
1014
Critical error:
Atmel CPU did
not confirm D5
actuation
Power off and on
call service
Replace
controller PCB
1015
Critical error:
12V power
failure
Power off and on
call service
Replace
controller board
or power supply
unit
1016
Critical error:
Atmel CPU and
C167 CPU AD
test failure
Power off and on
call service
Replace
controller Board
1017
Critical error:
Self-test: Atmel
CPU did not
perform AD test
Power off and on
call service
Replace
controller Board
1018
Critical error:
Atmel CPU clock
and C167 CPU
clock not
synchronized
Power off and on
call service
Replace CPU and
basis board
1019
Minor error:
RTC battery low
Confirm and
correct date/
time
call service
Replace RTC
battery