Service manual LIMAX
120
Service manual version 1.00
page 22
4.7
Power supply unit
The power supply unit delivers electrical energy for all parts of the LIMAX
120. The anode (1,
Fig. 4.5) and cathode (2) connectors are supplying the opto-mechanical unit. The keyswitch
connectors (5), the front panel connectors (6 and 7), the cooling unit connectors X1, X2, X3
(10, 11 and 12), the chiller and DI-water cooling power supply cables (8, 9) are located on the
front of the power supply unit. The Power supply unit is cooled with the help of the cooling
unit. For this purpose a water inlet connector is located on the top (3) and a water outlet
connector is located on the front end (4) of the power supply unit.
Fig. 4.5 Power supply unit
1
anode connector
7 front panel connector
2
cathode connector
8 chiller power supply
3
cooling water inlet
9 water pump supply connector
4
cooling water outlet
10 X1 connector
5
key switch connectors
11 X2 connector
6
front panel connector
12 X3 connector