4.9.2
Special voltages
Depending on nation different voltages are demand for devices. Therefore those versions are
featured with a toroid transformer (fig. 4.9c) which changes the usual 220-240V into 100 V,
110 V, 120 V or 127V.
Fig. 4.9c toroid transformer with a 5-pol terminal block
4.10
Footswitch
The footswitch is divided into two switches. The left SWAP-switch (1,Fig. 4.8) enables
switching between two predefined operating modes (A/B-mode). The right switch (2) activates
the laser and has two operating points. The first operating point activates the base airflow, the
base VAC-suction and any other auxiliary devices connected at the accessory port. The second
operating point activates the work flow, work suction and the laser if the LIMAX
120 is in
LASER READY mode. Releasing the switch the emission of laser light stops and the LIMAX
120 is set from LASER ACTIVE to LASER READY. The work flow and work suction changes to
base flow and base suction.
previous version
current version
Fig. 4.8 Footswitch
1
SWAP switch
2
Laser activating switch