142
Warning: EEPROM-memory is incorrect (power-off while
write access to shadow copy)
Measure(s): error cleared up by itself, no measure necessary
V1.16
143
Warning: EEPROM-content not consistent (power-off
while write access)
Measure(s): error cleared up by itself, no measure necessary
V1.16
Status-
Code
Error description
Since version
calibration errors
200
Critical error: data of blower calibration are invalid
Measure(s): restart blower calibration, readjust the blower
where applicable
v0.416
201
Minor error: Blower is not calibrated
Measure(s): readjust the blower where applicable and then
conduct the blower calibration
v0.420
9.6
Reset log file
Enter the menu item RESET LOGFILE by selecting the menu item and push down the rotary
switch (Fig. 9.6).