Service manual LIMAX
120
Service manual version 1.00
page 56
To start calibration select the menu item CONTINUE. The first calibration point will always
start with a current of 30A and will increase in 5A steps. Turn the laser into LASER READY,
activate the foot switch and note the measured laser power with a laser power meter. Then
select the menu item CONTINUE. You’ll get to the second calibration point with a current of
35A. Go on with the following points in the same way until you reach the last point with 90A
(13/13) or if laser power already amounts 120W some points before (see fig. 9.3c).
Fig. 9.3c Calibrate the laser diode
9.4
Cooling
Enter the menu item COOLING (Fig. 9.4).
Fig. 9.4 Enter the cooling menu