53
Critical error: watchdog-test failed
Measure(s): contact support
v0.138
Status-
Code
Error description
Since version
Flash/ROM errors
70
Critical error: bootloader failed
Measure(s): contact support
v0.307
71
Critical error: integrity of the flash/Rom is
incorrect/checksum error
Measure: reprogram firmware
v0.306
72
Critical error: checksum creation failed
Measure: reprogram firmware; if the error recurs use another
version of firmware or contact support
v0.306
Status-
Code
Error description
Since version
Hardware errors or systematic errors
100
Critical error: temperature sensor is not connected or
defect
Measure(s): Control connecting of the temperature sensor or
change it
v0.87
101
Severe error: overheat
Meausure(s): Cool off the fan
v0.87
102
Critical error: other (process) error of the temperature
sensor
Measure(s): contact support
v0.87
103
Critical error: clamping button while system start
Measure(s): Keypad can be activated when initializing has
finished
v0.87
104
Minor error: EEPROM-CRC was incorrect
v0.87
105
Critical error: clamping footswitch while system start
Measure(s): Footswitch can be activated when initializing has
finished
v0.113
106
Critical error: calibration of pressure sensor failed
Measure(s): Turn off device, let the fan run out, turn on
device; if the error occurs though the fan doesn’t work, it
indicates a defect pressure sensor
v0.128
107
Critical error: hardware error concerning measure of
blower current (configuration of VDR failed)
Measure(s): contact support
v0.129