![KLS Martin LIMAX 120 Service Manual Download Page 35](http://html1.mh-extra.com/html/kls-martin/limax-120/limax-120_service-manual_1979001035.webp)
Service manual LIMAX
120
Service manual version 1.00
page 32
5.3
Removal of the opto-mechanical unit
For replacement or servicing the opto-mechanical unit can be removed and placed in a secured
position. First shut down the LIMAX
®
120 laser unit and remove the backside cover (chapter
5.2). Remove four M4x10 socket head cap screws using a 3 mm hex screwdriver and move the
opto-mechanical unit backwards carefully without damaging the SMA-plus fibre connecting
socket. Rotate the opto-mechanical unit as seen in fig. 5.5 to place it in service position. When
replacing the opto-mechanical unit drain the DI-water cooling unit as described in chapter 0.
Remove all connectors before taking it out completely.
Fig. 5.5 Opto-mechanical unit in service position