PMX_MULTI
User’s Manual
75
Parallel Operation and Series Operation | Overview of Series Operation
Configuration for series operation
Setting the voltage and current
Set the voltage and current on CH2 (master). The voltage can be set up to double the voltage that is pos-
sible when operating independently.
Setting the overvoltage protection (OVP) and overcurrent protection (OCP)
Set the overvoltage protection (OVP) and overcurrent protection (OCP) on CH2 (master) (
Starting series operation
Turning on the POWER switch
Series operation cannot be used when the tracking function is enabled. Disable the tracking function.
1
Turn on the POWER switch of this product.
2
Press SHIFT+SER to switch to series operation mode.
The CH1 voltmeter shows “
SLAVE
,” and the CH2 display area shows “
SER
.”
Switching is possible only when the output is off.
3
Press SELECT on CH2.
CH2 is indicated with brackets.
4
Turn the VOLTAGE knob to set the voltage and the CURRENT knob to set the cur-
rent.
The CH2 display area shows the voltage and current.
5
Press OUTPUT on CH2 to turn output on.