PMX_MULTI
User’s Manual
21
Installation
Checking Whether the Power Is On or Off
Turning the POWER switch on
If the POWER switch is turned on for the first time after purchase, the product starts in the factory default
condition (
). Subsequent times that you turn the product on, it starts with the panel settings (excluding
the output on/off setting) that were in use immediately before the POWER switch was turned off.
You can use the CONFIG settings (CF01) to select the output state of the product when the POWER
switch is turned on
.
1
Check that the power cord is connected correctly.
2
Check that the output terminal cover is attached
.
When the product is shipped from the factory, the output terminal cover is not attached.
3
Turn the POWER switch on ( ).
All the indicators on the front panel turn on, and then the CH2 voltmeter and ammeter display the
following information in sequence: the rated voltage, rated current, and type, the firmware version
number, and then the build number. Each item is displayed for approximately 1 second.
After a few seconds, the product enters the operation standby state (the measured value is dis-
played).
WARNING
Risk of electric shock. Regardless of whether load cables are connected to the output
terminals, be sure to attach the output terminal cover before turning the POWER
switch on.
CAUTION
Risk of damage to load.
If the product is configured to turn on the output when the POWER switch is turned on in
CONFIG settings, set an appropriate OVP or OCP value before connecting a different load.
Firmware version
(Version 1.00 in this example)
Rated voltage, rated current, and type
(Example of PMX32-3TR)
Build number
(Build number BLD 0007 in this example)
CV
V
s
OFF
CC
A
s
ON
PARA
SERI
TRACKING 12
%
%
CV
V
s
OFF
CC
A
s
ON
PARA
SERI
TRACKING 12
%
%
CV
V
s
OFF
CC
A
s
ON
PARA
SERI
TRACKING 12
%
%