Protocol
The following table shows the settings that correspond to the RS232C protocol.
Underlined values are factory default settings.
Item
Setting
Baudrate
1 200 bps/ 2 400 bps/ 4 800 bps/ 9 600 bps/ 19 200 bps
Data
Fixed to 8 bit
Stop:
Fixed to 1 bit
Parity
Fixed to none
Flow
XFLOW/ NONE
RS232C communication
Use flow control for RS232C communication. DC (device control) codes are used as control
codes.
Transmission/reception may not work correctly through unilateral transmission.
Code
Function
ASCII code
DC1 (Xon)
Request to send
11H
DC3 (Xoff)
Transmission stop request 13H
Break signal
The break signal functions as a substitute for the IEEE488.1 dcl / sdc (Device Clear,
Selected Device Clear) message.
NOTE
The RS232C interface should be shifted remotely by the command. Use the "
"
SCPI command to set the RS232C interface to the remote state. Be sure to include this
command at the start of the program when you are performing remote programming.