6-2 Specifications
KJM6310
Trigger section
PLL clock regeneration section
Frequency response characteristics are valid for reference clock = 4.3 MHz.
*1
Complies with CD Reference: Compact Disc Reference Measuring Methods
Speci
fi
cation Guideline Ver.1.0 May 1999.
Output section
Symmetry follow-up
AUTO, AUTO+ OFFSET, or MANUAL
Trigger edge
RF
Rising edge, falling edge, or both
CLOCK
Rising edge or falling edge
Delay
Adjust the phase difference with the RF
signal by delaying the CLOCK signal
AUTO or MANUAL switchable
MANUAL adjustment range 0
°
to 360
°
Synchronizable signal
EFM signal that corresponds to a funda-
mental frequency of 4.1 MHz to 4.5
MHz.
Frequency response characteristics
(100 Hz reference, closed-loop charac-
teristics)
*1
1 kHz: 0.19
±
1.7 dB
5 kHz: –0.15
±
1.7 dB
10 kHz: –1.17
±
1.7 dB
20 kHz: –3.82
±
1.7 dB
25 kHz: –5.10
±
1.7 dB
Lockup time
Within 700 ms
Synchronizable jitter range
5 % to 17 %
Residual jitter
2 % or less
DC OUT
Output amplitude
(when open)
0.2 V/% and accuracy of
±
0.15 V
Output impedance
Approx. 600
Ω
JITTER
OUT
Output amplitude
(when open)
20 mV/%
Output impedance
Approx. 600
Ω
Summary of Contents for KJM6310
Page 1: ...KJM6310 Part No Z1 003 062 IA003161 Jan 2004 OPERATION MANUAL Time Interval Jitter Meter...
Page 8: ...VI Overvoltage category KJM6310...
Page 12: ...P 2 Preface KJM6310...
Page 50: ...5 4 Maintenance KJM6310...
Page 55: ......