Specifications | Basic specifications
Other functions
Item
Specifications
Trigger Function
Select external trigger (EXTERNAL) or internal trigger (INTERNAL).
INTERNAL
Measures at the sampling speed (FAST, MID, SLOW) interval.
EXTERNAL
Starts measurement with a SIGNAL I/O connector signal, *TRG, or the SNGL TRG key on
the front panel.
Trigger delay
0 to 9.999 s, OFF
Accuracy
±0.2 ms
Average function
The average count can be set between 2 and 99. OFF setting available.
Memory function
Saves up to 100 sets of measurement conditions.
key lock
Locks the key operation.
Zero adjustment
Zero adjustment of the voltmeter and resistance meter. OFF setting available. Zero point
clear function available.
Adjustment range
1000 digit
Measurement logging and
collective transfer function
Records up to 500 sets of measurement logs. Logs can be read collectively.
EOM function
Outputs an EOM signal from the SIGNAL I/O connector when a measurement is com-
pleted.
HOLD
When the trigger source is set to INTERNAL, the signal is turned on after a measurement is
completed until the next measurement starts. When the trigger source is set to EXTERNAL,
the signal is turned on after a measurement is completed until the next trigger is detected.
PULSE
Outputs a pulse when a measurement is completed.
Pulse width: 1 ms to 99 ms
Accuracy
±0.2 ms