User’s Manual
BIM1000 Series
Remote Control | Message Reference
Continuous measurement
The product performs continuous measurement and trigger wait even when :INITiate or READ? is not
sent.
Trigger wait
Switches from the idle state to the trigger wait state.
Trigger source
Selects whether to use the internal signal or an external signal for triggering.
Turning the delay on or off
Sets whether a delay will be inserted from when a trigger is detected until the measurement starts.
Delay
Sets the delay when :TRIGger:DELay:STATe is set to on.
Program header
:INITiate:CONTinuous
:INITiate:CONTinuous?
Value
{1|0|ON|OFF}
Response
{ON|OFF}
Note
This is automatically set to on when the power is turned on or when the product changes
to local mode.
Program header
:INITiate[:IMMediate]
Note
When continuous measurement (:INIT:CONT) is on, an execution error occurs.
Program header
:TRIGger:SOURce
:TRIGger:SOURce?
Value
{INTernal|EXTernal}
INTernal:
Uses the internal signal for triggering.
EXTernal:
Uses the SIGNAL I/O TRIG-IN signal, *TRG, or the SNGL TRG key on
the front panel for triggering.
Response
{INTERNAL|EXTERNAL}
Program header
:TRIGger:DELay:STATe
:TRIGger:DELay:STATe?
Value
{1|0|ON|OFF}
Response
{ON|OFF}
Program header
:TRIGger:DELay
:TRIGger:DELay?
Value/Response
<0 to 9.999 (NR2)>
Unit
Seconds