System Settings | General Settings
Current pulse output (CurrPulse)
This function prevents measurement errors caused by mutual interference of measurement currents, such
as when measurements are made using several of this product simultaneously. When Trigger source
(TRIG)
is set to EXTERNAL and CurrPulse is set to on, measurement current runs only during
measurement. Current interference can be prevented by offsetting the trigger inputs between several
instruments.
1
Press SYSTEM.
2
Press CurrPulse (F4) to select ON or OFF.
The value switches between ON and OFF each time that you press the F4 key.
This completes the setting.
External trigger
Measured current
1st scanner
External trigger
Measured current
2nd scanner