68
User’s Manual
BIM1000 Series
Remote Control | Message Reference
key lock
SIGNAL I/O lock
When the SIGNAL I/O lock is set to on, control through SIGNAL I/O will be disabled. This is automatically
turned off when the product is started.
Switches to local mode
During remote control, the product is in remote mode (no panel operations accepted), and a REMOTE
icon is shown in the display. To change to local mode (panel operations accepted), send the following
command. When the product enters local mode, the REMOTE icon disappears.
Saving and recalling measurement conditions
Saving measurement conditions
When you send a command with a designated memory number, measurement conditions is saved to the
specified memory number. Up to 100 entries can be saved to memory.
Recalling measurement conditions
When you send a command with a designated memory number, measurement conditions is recalled from
the specified memory number.
Program header
:SYSTem:KLOCk
:SYSTem:KLOCk?
Value
{1|0|ON|OFF}
Response
{ON|OFF}
Program header
:SYSTem:ELOCk
:SYSTem:ELOCk?
Value
{1|0|ON|OFF}
Response
{ON|OFF}
Program header
:SYSTem:LOCal
Program header
:SYSTem:SAVE
Value
<0 to 99 (NR1)>
Note
Note that there is a limit to the number of times the conditions can be saved (1 million).
Program header
:SYSTem:LOAD
Value
<0 to 99 (NR1)>
Note
If you specify an unsaved memory number, an execution error will occur.