![Kikusui BIM1000 Series User Manual Download Page 11](http://html.mh-extra.com/html/kikusui/bim1000-series/bim1000-series_user-manual_1966754011.webp)
BIM1000 Series
User’s Manual
11
Installation and Preparation
Connecting the DUT
Connecting the test leads to the product
1
Check that the POWER switch is turned off.
2
Connect the optional test leads
to the INPUT terminals.
The words “SENSE” and “SOURCE” are written on the test leads, which correspond to
the productʼs terminals. Connect the test leads to the appropriate terminals.
3
Turn the POWER switch ( ) on.
This completes the connections.
Examples of how to connect test leads to the DUT
Connect the positive and negative probes to the DUT’s positive and negative terminals, respectively.
This completes the procedure for connecting the test leads.
WARNING
Risk of electric shock.
• Do not make measurements exceeding the rated voltage of the test leads.
• When you connect or disconnect test leads from the INPUT connectors, make
sure that the tips of the test leads are not connected to the DUT.
The word
SENSE
The word
SOURCE
Example using a TL01-BIM
Summary of Contents for BIM1000 Series