Remote Control | Message Reference
Resistance judgment result
Voltage judgment result
Measurement logging and collective transfer function
When the measurement logging function is turned on, “LOG” is displayed in the upper right of the display.
With the function turned on, when you apply a SIGNAL I/O TRIG-IN signal or *TRG, a log of the measure-
ment will be saved in the product. The log saves up to 500 sets of the save number (STEP), resistance
measurement, and voltage measurement. Logs exceeding 500 sets are not saved. The log is cleared
when the power is turned off.
Log clear
Log count
Program header
:SENSe:JUDGment:RESistance:RESult?
Response
{UP|IN|LO|OFF|ERR}
UP:
The measurement exceeded the upper judgment limit.
IN:
The measurement is within the lower judgment limit and the upper judgment limit.
LO:
The measurement is less than the lower judgment limit.
OFF:
The judgment function is off.
ERR:
An error occurred.
Program header
:SENSe:JUDGment:VOLTage:RESult?
Response
{UP|IN|LO|OFF|ERR}
UP:
The measurement exceeded the upper judgment limit.
IN:
The measurement is within the lower judgment limit and the upper judgment limit.
LO:
The measurement is less than the lower judgment limit.
OFF:
The judgment function is off.
ERR:
An error occurred.
Program header
:LOG:STATe
:LOG:STATe?
Value
{1|0|ON|OFF}
Response
{ON|OFF}
Program header
:LOG:CLEAr
Program header
:LOG:COUNt?
Response
<1 to 500 (NR1)>