Remote Control | Message Reference
Voltage judgment function
Upper resistance judgment limit
Lower resistance judgment limit
Upper voltage judgment limit
Lower voltage judgment limit
Program header
:SENSe:JUDGment:VOLTage:STATe
:SENSe:JUDGment:VOLTage:STATe?
Value
{BOTH|OFF|UPPer|LOWer}
BOTH:
Turns on the upper and lower voltage limit judgment functions simultaneously
OFF:
Turns off the upper and lower voltage limit judgment functions simultaneously
UPPer:
Turns on only the upper voltage limit judgment
LOWer: Turns on only the lower voltage limit judgment
Response
{BOTH|OFF|UPPER|LOWER}
Note
If :SENSe:JUDGment:STATe is set to ON (1), the value changes to BOTH.
Program header
:SENSe:JUDGment:RESistance:UPPer
:SENSe:JUDGment:RESistance:UPPer?
Value/Response
<NR2>
Unit
Ω
Note
Set a value greater than the lower judgment limit.
Program header
:SENSe:JUDGment:RESistance:LOWer
:SENSe:JUDGment:RESistance:LOWer?
Value/Response
<NR2>
Unit
Ω
Note
Set a value less than the upper judgment limit.
Program header
:SENSe:JUDGment:VOLTage:UPPer
:SENSe:JUDGment:VOLTage:UPPer?
Value/Response
<NR2>
Unit
V
Note
Set a value greater than the lower judgment limit.
Program header
:SENSe:JUDGment:VOLTage:LOWer
:SENSe:JUDGment:VOLTage:LOWer?
Value/Response
<NR2>
Unit
V
Note
Set a value less than the upper judgment limit.