Calibration Procedures
1
Keysight U3606B Service Guide
97
CV output adjustment procedure — rear output terminals
Review the
“Input and Output Connections”
on page 26 sections before beginning this procedure.
1
Turn off the instrument and connect the rear panel
+
and
–
output terminals to
the
V
(red) and
LO
(black) input terminals.
2
Press
[Shift]
>
[EXT]
to enable remote sensing. When the U3606B is operating
in remote sensing mode, the
EXT
annunciator on the front panel is turned on.
3
Perform
through
of the
“CV output adjustment procedure — front
for the constant
voltage output adjustment items.
4
Repeat
through
for each output adjustment item shown in
.
NOTE
Ensure that the rear panel sense terminals (+S and –S) are shorted to the rear
panel output (+ and –) terminals. See
Chapter 3 of the U3606B User’s Guide,
“Remote sensing connections,” starting on page 116
for more information on
how to connect the load leads.
+S + + – – –S