1
Calibration Procedures
82
Keysight U3606B Service Guide
Frequency gain adjustment procedure
Review the
on page 25 and
on page 77 sections before beginning this procedure.
1
Press
[Hz ms %]
to enter the frequency gain calibration.
2
The display shows the uncalibrated value and the reference value of the
adjustment item.
3
Configure each adjustment item shown in
.
4
Use
or
to select the adjustment item.
5
Apply the input signal shown in the “Input” column of
6
Enter the actual applied input.
7
Press
to start the adjustment. The word CALib starts flashing to indicate
that the calibration is in progress.
– Successful completion of the adjustment is indicated by a short beep and
the display briefly showing PASS.
– An adjustment failure is indicated by a long beep, the display showing FAiL
along with a calibration error number. Correct the problem and repeat this
procedure.
8
Repeat
through
for each gain adjustment point shown in
.
9
Verify the frequency gain adjustments using the
NOTE
Always complete tests in the same order as shown in
Table 1-26
Frequency gain adjustment
Adjustment item
Input frequency
Input voltage (V
rms
)
1000.00 Hz, 1 V
rms
1 kHz
1 V
rms
OUT
SBY