Calibration Procedures
1
Keysight U3606B Service Guide
87
Capacitance gain adjustment procedure
Review the
on page 77 sections before beginning this procedure.
1
Press
[
]
to enter the capacitance gain calibration.
2
The display shows the uncalibrated value and the reference value of the
adjustment item.
3
Configure each adjustment item shown in
.
4
Use
or
to select the adjustment item.
5
Apply the input signal shown in the “Input” column of
6
Enter the actual applied input.
Table 1-29
AC current gain adjustment
Adjustment item
Input current
Input frequency
1.0000 mA, 1 kHz
1 mA
1 kHz
10.0000 mA, 1 kHz
10 mA
1 kHz
100.000 mA, 1 kHz
100 mA
1 kHz
1000.00 mA, 1 kHz
1000 mA
1 kHz
NOTE
If the zero offset adjustment procedure has been recently performed prior to the
capacitance gain calibration procedure, the adjustment item “OPEn” can be
omitted.
NOTE
Always complete tests in the same order as shown in
.
Summary of Contents for U3606B
Page 1: ...Keysight U3606B Multimeter DC Power Supply Service Guide ...
Page 8: ...8 Keysight U3606B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 12: ...12 Keysight U3606B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 14: ...14 Keysight U3606B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...