1
Calibration Procedures
92
Keysight U3606B Service Guide
CV output adjustment procedure — front output terminals
Review the
“Input and Output Connections”
on page 26 sections before beginning this procedure.
1
Turn off the instrument and connect the front panel
(red) and (black)
output terminals to the
V
(red) and
LO
(black) input terminals.
2
Turn on the instrument. Press
[Voltage]
to enter the constant voltage output
calibration.
3
The display shows the uncalibrated value and the reference value of the
adjustment item.
4
Configure each adjustment item shown in
.
LOAd1
[a]
100 mA with additional 300
Ω
(3 W) fixed resistor
LOAd2
[a][b]
6 mA with additional 1000
Ω
fixed resistor
LOAd3
1 A with additional 30
Ω
(30 W) fixed resistor
LOAd4
3 A with additional 2.67
Ω
(24 W) fixed resistor
[a] See
for information on how to connect the LOAd adjustment item.
[b] LO RES load.
Table 1-32
Constant current output adjustment (continued)
Adjustment item
Output current