Calibration Procedures
1
Keysight U3606B Service Guide
91
6
Press
to start the adjustment. The word CALib starts flashing to indicate
that the calibration is in progress.
– Successful completion of the adjustment is indicated by a short beep and
the display briefly showing PASS.
– An adjustment failure is indicated by a long beep, the display showing FAiL
along with a calibration error number. Correct the problem and repeat this
procedure.
7
Repeat
through
for each output adjustment item shown in
.
For adjustment item “LOAd1”, “LOAd2”, “LOAd3”, and “LOAd4”, connect an
additional fixed resistor across the front panel
(red) output terminal and the
I
(red) terminal according to the fixed resistors value on
. Leave the
(black) terminal and
LO
(black) terminal connected.
8
Verify the constant current adjustments using the
Table 1-32
Constant current output adjustment
Adjustment item
Output current
OUt-L
AUTO calibration
OUt-H
AUTO calibration
OUT
SBY
Fixed resistor