1
Calibration Procedures
90
Keysight U3606B Service Guide
CC output adjustment procedure — front output terminals
Review the
“Input and Output Connections”
on page 26 sections before beginning this procedure.
1
Turn off the instrument and connect the front panel
(red) and (black)
output terminals to the
I
(red) and
LO
(black) input terminals.
2
Turn on the instrument. Press
Current
to enter the constant current output
calibration.
3
The display shows the uncalibrated value and the reference value of the
adjustment item.
4
Configure each adjustment item shown in
.
5
Use
or
to select the adjustment item.
NOTE
Always complete tests in the same order as shown in