1
Calibration Procedures
26
Keysight U3606B Service Guide
Input and Output Connections
Test connections to the instrument are best accomplished by using a dual banana
plug with a copper wire short between the two terminals for low-thermal offset
measurement. Shielded, twisted-pair PTFE interconnecting cables of minimum
length are recommended between the calibrator and the U3606B. Cable shields
should be earth ground referenced. This configuration is recommended for
optimal noises and settling time performance during calibration.
Summary of Contents for U3606B
Page 1: ...Keysight U3606B Multimeter DC Power Supply Service Guide ...
Page 8: ...8 Keysight U3606B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 12: ...12 Keysight U3606B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 14: ...14 Keysight U3606B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...