1
Calibration Procedures
56
Keysight U3606B Service Guide
CC programming and readback accuracy
This test verifies that the constant current programming and readback accuracy
are within published specifications. The accuracy of the current monitoring
resistor must be 0.1% or better.
1
Turn off the instrument and connect a 0.01
Ω
current shunt monitoring resistor
across the front panel
(red) and (black) output terminals and a digital
multimeter across the current shunt monitoring resistor (see
setup for constant current programming and readback accuracy verification,”
2
Turn on the instrument. Press
[Current]
to select the constant current mode.
Ensure that the appropriate range is selected.
3
Output each of the full rated value in the order shown below. Divide the
voltage drop (reading on the digital multimeter) across the current monitoring
resistor by its resistance. Record and compare the measurement results to the
appropriate test limits shown in
.
4
Repeat
for the S2 (8 V/3 A) and S1m (100 mA/30 V) ranges.
Table 1-18
Constant current programming and readback accuracy
verification test
Output current
Range (parameter)
Error from nominal 1-year
0 A
S1 (1 A)
±0.003 A
1 A
±0.0045 A
0 A
S2 (3 A)
±0.003 A
3 A
±0.0075 A
0 A
S1m (100 mA)
±0.00015 A
0.1 A
±0.0002 A
Summary of Contents for U3606B
Page 1: ...Keysight U3606B Multimeter DC Power Supply Service Guide ...
Page 8: ...8 Keysight U3606B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 12: ...12 Keysight U3606B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 14: ...14 Keysight U3606B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...