1
Calibration Procedures
96
Keysight U3606B Service Guide
For adjustment item “LOAd1”, “LOAd2”, “LOAd3”, and “LOAd4”, connect an
additional fixed resistor across the front panel + rear output terminal and the
I
(red) terminal according to the fixed resistors value on
. Leave the –
rear terminal and
LO
(black) terminal connected.
5
Verify the constant current adjustments using the
NOTE
Do not remove the short bar between the rear panel sense terminals (+S and –S)
are the rear panel output (+ and –) terminals. See
Chapter 3 of the U3606B
User’s Guide, “Remote sensing connections,” starting on page 116
for more
information on how to connect the load leads.
+S + + – – –S