To replace the current input fuse
The current input terminal is fuse protected. The fuse for the current input
terminal is located on the front panel. The fuse is a 3.15 A/500 V F/B 6.3
×
32 mm
fuse, Keysight part number A02-62-25657-1U. If you are certain that the fuse is
faulty, replace it with one of the same size and rating.
1
Disconnect all input terminal connections. Place a flat blade screwdriver on
the fuse holder assembly as shown below and push in.
2
Turn the screwdriver anti-clockwise to remove the fuse holder assembly from
the front panel.