1
Calibration Procedures
62
Keysight U3606B Service Guide
Additional Verification Tests
Optional capacitance gain verification test
1
Connect the calibrator to the front panel (red) and
LO
(black) input
terminals (see
Figure 1-3, “Test setup for capacitance gain verification,”
2
Press [
] to select the capacitance function.
a
For ranges lower than 1.000 µF, please perform Null at the
device-under-test (DUT) before recording the final measurement. This will
compensate the capacitance errors present in the test leads and terminals
of the Fluke 5520A calibrator.
b
Output 190 pF (equivalent to 0.19 nF) from the Fluke 5520A Calibrator
(
NORMAL HI+LO
) to the DUT terminals (
V
+
LO
). The lowest capacitance
value that can be sourced from the Fluke 5520A calibrator is 190 pF.
c
Select the desired range to perform measurement; as an example, for the
10.00 nF range, set measurements to continuous auto trigger, and then
press
[Null]
to perform the Null offset.
d
Press
[Null]
again to edit the Null offset value.
e
Deduct the Null value displayed on the DUT's front panel with 190 pF
(equivalent to 0.19 nF) using the arrow keys and press
[Shift]
>
[Save]
to
save the new Null offset value.
f
Perform the subsequent measurements in the desired active range as
with the new edited Null offset value saved in the DUT.
3
for any change of ranges that is lower
than 1.000 µF.
Table 1-22
Optional capacitance gain verification test
Input capacitance
Range
Error from nominal 1-year
1 nF
1 nF
±0.028 nF
10 nF
10 nF
±0.15 nF
100 nF
100 nF
±1.5 nF
1
μ
F
1
μ
F
±15 nF
Summary of Contents for U3606B
Page 1: ...Keysight U3606B Multimeter DC Power Supply Service Guide ...
Page 8: ...8 Keysight U3606B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 12: ...12 Keysight U3606B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 14: ...14 Keysight U3606B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...