Calibration Procedures
1
Keysight U3606B Service Guide
93
5
Use
or
to select the adjustment item.
6
Press
to start the adjustment. The word CALib starts flashing to indicate
that the calibration is in progress.
– Successful completion of the adjustment is indicated by a short beep and
the display briefly showing PASS.
– An adjustment failure is indicated by a long beep, the display showing FAiL
along with a calibration error number. Correct the problem and repeat this
procedure.
7
Repeat
through
for each output adjustment item shown in
.
For adjustment item “LOAd1”, “LOAd2”, “LOAd3”, and “LOAd4”, connect an
additional fixed resistor the front panel
(red) and (black) output
terminals according to the fixed resistors value on
. Leave the
connections from the output terminals (
and ) to the input terminals
(
V
and
LO
) intact.
NOTE
Always complete tests in the same order as shown in
.
OUT
SBY