Calibrator Output Operations
2
Keysight U1401B User’s and Service Guide
51
Defining the autoscan parameters in the memory
Press and hold
MODE
for more than one second to enter the autoscan adjustment
mode. A total of 16 steps with individually definable time interval and amplitude
are available.
When the instrument is in the autoscan adjustment mode, the secondary display
shows the amplitude. The first two digits of the primary display are used to
indicate which step is being adjusted. The last two digits of the primary display
are used to indicate time interval.
1
Press
MODE
to cycle through step, time interval, and amplitude adjustments.
The digit to be adjusted will flash on the display.
– For amplitude adjustment, press
and
to select the digit to be
adjusted, and then press
and
to adjust the value of the selected
digit. The amplitude can be set to any value within the selected output
range (±1.5 V or ±15 V for constant voltage output, ±25 mA for constant
current output).
– For time interval adjustment, press
and
to select the digit to be
adjusted, and then press
and
to adjust the value of the selected
digit. The time interval can be set within the range of 0 to 99 seconds.
– Press
for more than one second to directly reset the time interval and
amplitude of the present step to zero.