Performance Tests and Calibration
7
Keysight U1401B User’s and Service Guide
153
Output calibration
1
Set the slide switch to the
M/S
position.
2
Allow the instrument to warm up for 10 minutes before performing calibration.
3
To enter the calibration mode, press
and
for more than one
second. The primary display will indicate “CHEEP”.
4
Press
to enter the input calibration mode.
5
Turn the rotary switch to any one of the “Current Input/Voltage Output”
positions, and press
SHIFT
for more than one second to enter the output
calibration mode.
CAL-0 & CAL-1
In the output calibration mode, the primary and secondary display will show
“CAL-0” and “-rdy-” respectively.
Connect the output terminals to a multimeter (refer to
the recommended test equipment).
– CAL-0:
1
Press
OUTPUT
. The primary and secondary displays show “CAL-0” and
“00000” respectively.
2
Wait until the instrument reading becomes stable, then record the value.
– CAL-1:
1
Press
MODE
. The primary and secondary displays show “CAL-1” and “-rdy”
respectively.
2
Press
OUTPUT
. The primary and secondary displays show “CAL-1” and
“00000” respectively.
3
Press
or
to adjust the output voltage until the reading on the meter
is the same as the
CAL-0
value recorded above.
4
Press
MODE
to finish the
CAL-0
and
CAL-1
calibration.
After finishing the
CAL-0
and
CAL-1
calibration procedures, the instrument will
automatically enter the 1.5 V output calibration mode.