Chapter 4: Calibrating and Adjusting
To run the self calibration
70
Figure 4-1
Calibration Dialog
3
Connect an adapter to each end of the cable.
4
Select Start, then follow the instructions on the screen.
You will be asked first to disconnect all channels and the Aux Output (located on the
rear panel of the oscilloscope), then to connect Aux Out to each channel and the
auxiliary trigger in turn. After you complete each cable change, select OK in the dialog
box to continue the calibration.
The oscilloscope displays a Passed/Failed message as each calibration routine is
completed.
5
Select Close.
Clear Cal Memory Protect to Run self calibration
You cannot run self calibration if this box is checked.
If calibration fails
Go to chapter 5 “Troubleshooting.”
Clear this check box
before starting
calibration
Select here to start
calibration