Chapter 4: Calibrating and Adjusting
To run the self calibration
69
To run the self calibration
Equipment Required
1
Select Calibration from the Utilities menu.
2
If the Cal Memory Protect box in the Calibration dialog is checked, select the
check box to clear the check mark.
Equipment
Critical Specifications
Recommended Model/Part
Adapters (2)
BNC (f) to BNC (m) right angle
Keysight 1250-0076
Cable
BNC 50
Ω
Keysight
8120-1839
Let the Oscilloscope Warm Up Before Running the Self Calibration
The self calibration should only be done after the oscilloscope has run for 30 minutes at ambient temperature
with the covers installed.
Calibration time
It will take approximately 45 minutes to run the self calibration on the oscilloscope, including the time required
to change cables from channel to channel.