Chapter 3: Testing Performance
To verify threshold accuracy
54
5
Set up the Acquisition system as follows:
a
Select Setup from the menu bar, and then select Acquisition...
b
Set the analog memory depth.
Select Manual in the Analog Memory Depth pane.
Set the number of points to 16.
Note: Minimizing acquisition memory depth allows the oscilloscope to display
changing inputs more quickly and this reduces the chance for making an error
in this procedure.
Close the Acquisition Setup dialog.
c
Select Setup from the menu bar, and then select Trigger.... Ensure that the trigger
mode is set to Edge and Sweep is set to Auto.