Chapter 6: Replacing Assemblies
To remove and replace the internal digital input cable (MSO models only)
106
To remove and replace the internal digital input cable (MSO
models only)
Use this procedure to remove and replace the internal digital cable. When necessary,
refer to other removal procedures. The graphics in this chapter are representative of
the oscilloscope at the time of this printing. Your unit may look different.
1
Disconnect the power cable and remove the bottom sleeve.
2
Remove the AutoProbe board assembly from the front frame (see page 104).
3
Use a small screwdriver to remove the spring clip from the internal digital cable
connector then disconnect the cable from the acquisition board.
4
Clip the tie wraps holding the digital cable and shield to the side of the chassis.
5
Using a Torx T10 driver remove the screw holding the internal digital cable to
the front panel.
Figure 6-11
Removing the internal digital cable
6
Pull the internal digital cable out through the rear of the front panel.
7
To replace the internal digital cable, reverse the above procedure.
Shield
T10 torx
Internal
digital
cable
Acquisition
board
Tie wrap
Spring clip
Tie wrap
Summary of Contents for Infiniium 8000 Series
Page 6: ...6...
Page 31: ...Chapter 2 Preparing for Use Digital probe lead set 31 Data Sheets Demonstrations Catalogs...
Page 66: ...Chapter 3 Testing Performance Performance Test Record 66...
Page 140: ...Chapter 7 Replaceable Parts Exploded Views 140 Exploded Views Front Frame and Front Panel...
Page 141: ...Chapter 7 Replaceable Parts Exploded Views 141 Fan and Acquisition Assembly...
Page 143: ...Chapter 7 Replaceable Parts Exploded Views 143 Power Supply and PC Motherboard With Option 017...
Page 144: ...Chapter 7 Replaceable Parts Exploded Views 144 Sleeve and Accessory Pouch...
Page 150: ...Chapter 8 Theory of Operation 150 Figure 8 1 Oscilloscope Block Diagram...
Page 158: ...158...
Page 160: ...160 Index...