Chapter 3: Testing Performance
To test single cursor voltage measurement accuracy with offset
45
8
Select Vavg from the Voltage submenu of the Measure menu. Ensure that Channel
1 is selected in the "Enter Measurement Info" dialog and select OK.
#1 should be +- 382mV 8.368V 9.132V
#2 should be +- 77mV 1.67V 1.82V
Figure 3-5
Source Selection for Vavg Measurement
9
Select Channel 1 from the Setup menu.
10
Set the sensitivity for Channel 1 to the volts/div value from the first row of
the table in step 7. Set the Offset control to the offset value from the first row
of the same table. Select Close.
Sensitivity Channel
Vin
Tolerance
Limits
(Volts/div)
Offset
(Supply)
Minimum
Maximum
5 V
17.5 V
35 V
+/- 1.60 V
33.40 V
36.60 V
2 V
7 V
14 V
+/- 641 mV
13.36 V
14.64 V
1.25 V
4.375 V
8.75 V
+/- 401 mV
8.349 V
9.151 V
500 mV
1.75 V
3.5 V
+/- 161 mV
3.339 V
3.661 V
250 mV
875 mV
1.75 V
+/- 80.9 mV
1.669 V
1.831 V
100 mV
350 mV
700 mV
+/- 33.0 mV
667.0 mV
733.0 mV
50 mV
175 mV
350 mV
+/- 17.0 mV
333.0 mV
367.0 mV
20 mV
70 mV
140 mV
+/- 7.40 mV
132.6 mV
147.4 mV
10 mV
35 mV
70 mV
+/- 4.20 mV
65.80 mV
74.20 mV
5 mV
17.5 mV
35 mV
+/- 2.60 mV
32.40 mV
37.60 mV
Select Channel 1 as the
source for the Vavg
measurement
Summary of Contents for Infiniium 8000 Series
Page 6: ...6...
Page 31: ...Chapter 2 Preparing for Use Digital probe lead set 31 Data Sheets Demonstrations Catalogs...
Page 66: ...Chapter 3 Testing Performance Performance Test Record 66...
Page 140: ...Chapter 7 Replaceable Parts Exploded Views 140 Exploded Views Front Frame and Front Panel...
Page 141: ...Chapter 7 Replaceable Parts Exploded Views 141 Fan and Acquisition Assembly...
Page 143: ...Chapter 7 Replaceable Parts Exploded Views 143 Power Supply and PC Motherboard With Option 017...
Page 144: ...Chapter 7 Replaceable Parts Exploded Views 144 Sleeve and Accessory Pouch...
Page 150: ...Chapter 8 Theory of Operation 150 Figure 8 1 Oscilloscope Block Diagram...
Page 158: ...158...
Page 160: ...160 Index...