Chapter 3: Testing Performance
To verify threshold accuracy
52
f
Select Measure from the menu bar and then select Markers...
g
Set the mode to Track Waveforms.
h
Set the position of both markers to center screen by selecting the 0 buttons for
each marker.
i
Close the Marker Setup window.
j
Display the states of the digital bits at the markers in both hexadecimal and
binary forms by selecting the Logic tab in the marker display area at the bottom
of the LCD display.
Note: The data in the left column shows the states of the digital inputs at the
Cursor A position and the data in the right column shows the states of the digital
inputs at the Cursor B position.
k
Set the oscilloscope digital threshold to the value being tested (e.g. +5V, 0V or -
5V) as follows:
From the menu bar, select Setup, then select Digital..., then select the Threshold
tab. Make the following settings for both pods:
Set the Logic Family to User Defined.
Set the Threshold for both pods to the value being tested. Start with the first
Select here
Select here
Select the
logic tab
Summary of Contents for Infiniium 8000 Series
Page 6: ...6...
Page 31: ...Chapter 2 Preparing for Use Digital probe lead set 31 Data Sheets Demonstrations Catalogs...
Page 66: ...Chapter 3 Testing Performance Performance Test Record 66...
Page 140: ...Chapter 7 Replaceable Parts Exploded Views 140 Exploded Views Front Frame and Front Panel...
Page 141: ...Chapter 7 Replaceable Parts Exploded Views 141 Fan and Acquisition Assembly...
Page 143: ...Chapter 7 Replaceable Parts Exploded Views 143 Power Supply and PC Motherboard With Option 017...
Page 144: ...Chapter 7 Replaceable Parts Exploded Views 144 Sleeve and Accessory Pouch...
Page 150: ...Chapter 8 Theory of Operation 150 Figure 8 1 Oscilloscope Block Diagram...
Page 158: ...158...
Page 160: ...160 Index...